Anaesthesia Machine Atlas Anesthesia Briefs

Interactive teaching atlas · Chapter 1 · Machine

Anaesthesia Machine Atlas

Explore an anaesthesia workstation part by part: the path gas takes from the wall outlet to the alveolus, the physical laws that govern it, the bellows ventilator and the patient monitor. Rotate the model, tap the numbered parts and adjust the flows.

18 parts10 laws of physics8 equipment models4 cases26 machine faults7 clinical events7 questions
Drag: rotate · Move: arrow pad, Shift + drag, right-click · Full screen: two fingers or arrow keys · Ctrl + wheel: zoom

    Chapter 2 · Gas path

    From wall to alveolus: three pressure zones

    The machine lowers gas pressure step by step. The 150 bar in the cylinder falls to around 20 cmH₂O in the patient's airway, a difference of roughly 7500-fold. At each step a different physical law and a different safety feature come into play.

    High pressure

    150bar
    • Cylinder, hanger yoke, pin index system
    • Cylinder pressure gauge (Bourdon)
    • First-stage regulator: output ≈ 3–4 bar
    • N₂O: liquid + vapour at 20 °C, constant ≈ 51 bar

    Intermediate pressure

    ≈ 4bar
    • Pipeline inlets and non-return valves
    • O₂ flush (35–75 L/min), O₂ supply failure alarm
    • Hypoxic guard, ventilator driving gas
    • Flow control needle valves

    Low pressure

    < 0.1bar above ambient
    • Flowmeter tubes (Thorpe)
    • Vaporisers and interlock
    • Common gas outlet → circle system
    • Most frequent leak site: negative-pressure test

      Chapter 3 · Ventilator

      Bellows ventilator and the equation of motion

      Most anaesthesia ventilators are double-circuit: driving gas squeezes the bellows from outside while patient gas stays inside it. The bellows below moves in step with the lung in the simulator. Try the six modes (VCV, PCV, PRVC, SIMV, PSV, CPAP), increase the patient's breathing effort and watch the triggering.

      Blue mist: driving gas filling the housing during inspiration

      Patient type

      Lung presets

      Ventilation mode

      450mL
      12/min
      1:2
      5cmH₂O
      50mL/cmH₂O
      10cmH₂O·s/L
      170cm

      Patient's breathing effort

      0cmH₂O
      14/min
      0,9s

      Trigger and cycling

      2L/min

      0 = patient under neuromuscular block, not breathing. As effort increases the patient triggers the ventilator (▲ above the waveform).

      Pressure–volume loop. The slope shows compliance, the width of the loop shows the resistive pressure.

      What is the screen telling you?

      The model is a single-compartment linear lung solved in real time in 2.5 ms steps. VCV uses square-wave flow with a 10% inspiratory pause; triggering uses an adjustable flow threshold (adult default 2 L/min). Labels above the waveform show the breath type: VC, PC, PS, SP; ▲ patient-triggered, BACK-UP apnoea ventilation. Circuit compliance and fresh gas coupling are ignored.

      Patient–ventilator asynchrony

      Mismatches between the patient's effort and the ventilator's breaths. Markers on the waveform screen: ✕ IE ineffective triggering, DT double triggering, FS flow starvation, DC delayed cycling, AP flow not returning to zero at end-expiration (auto-PEEP).

      Last 60 seconds

      Chapter 4 · Monitor

      Monitor: reading the waveforms

      A real-time physiology model runs behind the monitor: circuit gas, alveolar O₂ and CO₂ change over minutes. If FiO₂ or ventilation fails, SpO₂ falls slowly; arrhythmia, bradycardia and arrest develop only with prolonged severe hypoxaemia. NIBP measures blood pressure intermittently; insert an arterial line for continuous monitoring. Start a machine fault or clinical event from the catalogue below, read the findings and choose the correct intervention.

      Monitor and simulation

      Simulation speed
      Patient
      Altitude
      NIBP

      Abbreviations: HR heart rate · PR pulse rate · RR respiratory rate · NIBP non-invasive blood pressure · ART invasive arterial pressure · MAC minimum alveolar concentration.

      NIBP only shows the pressure at the moment it measures; only an arterial line continuously captures a fall between two readings.

      Patient: what the monitor cannot see

      Case mode

      Hidden events arrive in sequence during an operation. Read the findings and intervene from the event card above. At the end you get a debriefing report and a score.

      Event catalogue

      Choose an event. In hidden fault mode the name and description are hidden; you have to diagnose from the findings.

      ECG

      The electrical activity of the heart is measured as a potential difference of a few millivolts between surface electrodes. In theatre, lead II (rhythm, P waves) and V5 (left ventricular ischaemia) are most often monitored together. Diathermy interference and filter settings can artificially alter the ST segment: diagnostic mode (0.05–150 Hz) and monitoring mode (0.5–40 Hz) are not the same.

      Pulse oximetry and plethysmography

      Oxyhaemoglobin absorbs more infrared, deoxyhaemoglobin more red light. The device isolates arterial blood by using only the pulsatile (AC) component. Carboxyhaemoglobin makes SpO₂ falsely high, and methaemoglobin pulls it towards ≈ 85%. Respiratory variation of the plethysmograph can indicate hypovolaemia.

      Capnography

      A side-stream capnometer draws 50–200 mL/min of sample from the circuit; a mainstream sensor fits directly on the Y-piece. Normally EtCO₂ is 2–5 mmHg lower than PaCO₂; the gap widens with dead space.

      I II III 0 α β EtCO₂ dead space alveolar plateau inspiration

      NIBP, temperature and gas analysis

      Oscillometric NIBP tracks arterial wall oscillations as the cuff deflates; the pressure at maximum oscillation is mean arterial pressure, and systolic and diastolic values are calculated by algorithm. Cuff width should be ≈ 40% of arm circumference. O₂ is measured with a paramagnetic or galvanic cell and anaesthetic agents by infrared absorption; the MAC value is the sum of the agents' contributions.

      Chapter 5 · Safety

      Fail-safe design

      Most of the safety systems on an anaesthesia machine are a response to an accident in the past. None of them is sufficient on its own: the last line of defence against a hypoxic mixture is the O₂ analyser in the inspiratory limb and the clinician watching it.

      Pin index safety system

      The holes in the cylinder valve and the pins on the yoke are specific to each gas, so the wrong cylinder cannot be fitted. A missing pin or two washers stacked together defeats the system.

      DISS / NIST connectors

      Pipeline hoses connect with gas-specific diameters and thread profiles. Wall outlets also use gas-specific quick connectors.

      Hypoxic guard

      The N₂O and O₂ flow knobs are linked mechanically (chain and sprockets) or pneumatically. The Datex-Ohmeda Link-25 system limits the N₂O:O₂ ratio to 3:1, so the fresh gas contains at least 25% O₂; this does not guarantee the FiO₂ the patient inspires. Air or a third gas is outside this protection.

      Oxygen failure protection device

      When O₂ pressure falls (fail-safe valve) the flow of N₂O and other gases is cut off or reduced in proportion; current standards require an alarm within 5 seconds once pressure falls below the manufacturer's threshold (usually ≈ 2 bar / 30 psig) (older mechanical systems sounded for at least 7 seconds). This system monitors pressure, not flow or concentration.

      O₂ flowmeter last

      The O₂ tube is positioned nearest the common manifold (last, downstream). If an upstream tube leaks, the gas lost is N₂O or air, not O₂.

      Vaporiser interlock and agent-specific filling

      The interlock allows only one vaporiser to be turned on at a time. Keyed filling ports prevent filling with the wrong agent; if an agent with a high SVP is put into the vaporiser of an agent with a lower SVP, an overdose results.

      Hands-on machine check

      Find the faulty machine

      A hidden fault is randomly assigned to the machine (or the machine is fine). Carry out the tests in ASA 2008 order, mark each step "Pass" or "Fail", then choose the faulty component.

      Ready

        Pin index positions

        Positions of the yoke pins by gas (ISO 407). Choose a gas.

        Machine checklist

        Abridged from the ASA 2008 recommendations. Your ticks are stored only in this browser. The manufacturer's checkout procedure always takes priority.

          Chapter 6 · Test

          Test yourself

          Each question builds on a concept from the earlier chapters. When you tap an option, the correct answer and a short explanation appear.

          0 / 7 correct